LEICA RM2255 1492255UL01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-08-24 for LEICA RM2255 1492255UL01 manufactured by Leica Biosystems Nussloch Gmbh.

Event Text Entries

[52847139] This incident was user related. While installing the object holder on new microtome rm2255 the leica biosystems employee didn't realize the customer had placed a blade in the blade holder. The blade holder moved and he cut his finger. He required stitches. The instructions for use includes the following safety instructions: "when working with microtomes, personal safety precautions must always be taken. It is mandatory to wear work safety shoes, safety gloves, a mask and safety goggles. Take care when handling microtome knives and disposable blades. The cutting edge is extremely sharp and can cause serious injuries! "
Patient Sequence No: 1, Text Type: N, H10


[52847140] A leica biosystems employee was installing a new microtome rm2255 on customer site and caused a cutting damage on his finger. Medical treatment was necessary.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8010478-2016-00005
MDR Report Key5899674
Report SourceHEALTH PROFESSIONAL
Date Received2016-08-24
Date of Report2016-08-10
Date of Event2016-08-10
Date Mfgr Received2016-08-10
Device Manufacturer Date2016-05-21
Date Added to Maude2016-08-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactROBERT GROPP
Manufacturer StreetHEIDELBERGERSTR. 17-19
Manufacturer CityNUSSLOCH, 69226
Manufacturer CountryGM
Manufacturer Postal69226
Manufacturer Phone9622414334
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLEICA RM2255
Generic NameMICROTOM, PRODUCT CODE: IDO
Product CodeIDO
Date Received2016-08-24
Model Number1492255UL01
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS NUSSLOCH GMBH
Manufacturer AddressHEIDELBERGERSTR. 17 NUSSLOCH, 69226 GM 69226


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-08-24

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.