HARMONY IQ SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-08-31 for HARMONY IQ SYSTEM manufactured by Vts Medical Systems.

Event Text Entries

[54401229] A steris service technician inspected the system and found that the cause of the reported event is due to polling of the harmony iq system to the monitor. Polling is when the harmony iq system "pings" the monitor to obtain a status. Investigation of this event is currently in process. A follow up report will be submitted when additional information becomes available.
Patient Sequence No: 1, Text Type: N, H10


[54401230] The user facility reported that during a patient procedure their monitor powered down. No injuries or procedural delays or cancellations were reported.
Patient Sequence No: 1, Text Type: D, B5


[63647123] The steris service technician made proper repairs to the harmony iq system by removing the polling and confirmed the system to be operating properly.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1000404456-2016-00006
MDR Report Key5919170
Date Received2016-08-31
Date of Report2016-08-31
Date of Event2016-08-02
Date Mfgr Received2016-08-02
Date Added to Maude2016-08-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1VTS MEDICAL SYSTEMS
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHARMONY IQ SYSTEM
Generic NameCUSTOMIZED RACK
Product CodeKQM
Date Received2016-08-31
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVTS MEDICAL SYSTEMS
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2016-08-31

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.