C-MAC 8402X

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-08-31 for C-MAC 8402X manufactured by Karl Storz Gmbh & Co. Kg Tuttlingen, Germany.

Event Text Entries

[53665131] We evaluated the electronic module and found that some of the male pins on mating connector of blade were bent significantly to cause misalignment with the sockets on this module. The resulting impact caused one of the sockets to split and partially become pushed back into the plastic connector body. As a result, the integrity of the electrical connection was compromised. This fault was ultimately caused by improper care and handling of the equipment.
Patient Sequence No: 1, Text Type: N, H10


[53665132] Allegedly, doctor was performing an intubation on a (b)(6) baby; the light source failed once blade was placed in patient. When the handle was turned, image became intermittent. The staff didn't have a backup electronic module, so they swapped out the blade to see whether that would solve the problem. They used a competitor's glidescope instead to complete the procedure. Per the hospital, the patient expired, which they believe was related to patient condition, but the delay in switching between blades didn't likely help the situation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610617-2016-00131
MDR Report Key5919593
Date Received2016-08-31
Date of Report2016-08-03
Date of Event2016-08-02
Date Mfgr Received2016-08-03
Date Added to Maude2016-08-31
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag0
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. SUSIE CHEN
Manufacturer Street2151 E. GRAND AVENUE
Manufacturer CityEL SEGUNDO CA 902455017
Manufacturer CountryUS
Manufacturer Postal902455017
Manufacturer Phone4242188201
Manufacturer G1KARL STORZ GMBH & CO. KG TUTTLINGEN, GERMANY
Manufacturer StreetMITTLESTRASSE 8, 78503
Manufacturer CityTUTTLINGEN, GERMANY,
Manufacturer CountryGM
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameC-MAC
Generic NameELECTRONIC IMAGING MODULE
Product CodeOGH
Date Received2016-08-31
Returned To Mfg2016-08-08
Model Number8402X
Catalog Number8402X
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No0
Device Event Key0
ManufacturerKARL STORZ GMBH & CO. KG TUTTLINGEN, GERMANY
Manufacturer AddressMITTLESTRASSE 8, 78503 TUTTLINGEN, GERMANY, GM


Patients

Patient NumberTreatmentOutcomeDate
101. Death 2016-08-31