VITEK? 2 AST-YS07 TEST KIT 414967

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-09-01 for VITEK? 2 AST-YS07 TEST KIT 414967 manufactured by Biomerieux, Inc.

Event Text Entries

[53729759] Device not returned to manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[53729760] A customer in the (b)(6) reported the occurrence of discrepant ast results (intermediate and/or susceptible) for a (b)(6) survey sample (b)(6) (candida albicans) in association with the vitek 2 ast-ys07 test kit. Fluconazole i/s (mic = 4mg/l). Voriconazole s (mic <= 0. 12mg/l). The expected results were resistant. Fluconazole i/s (mic > 64mg/l). Voriconazole s (mic > 8mg/l). Note: in the final report from (b)(6), fluconazole and voriconazole were not graded since the "levels of concordance with fluconazole (63%) and voriconazole (59%) did not reach the requisite 80% of participants. " the customer repeated testing using the vitek 2 ast-ys07 test kit and obtained susceptible results of <=1mg/l for fluconazole and <=0. 12mg/l for voriconazole. It is unknown if any alternate method was employed to confirm the expected result. There is no indication or report from the hospital to biomerieux that the discrepant result led to any adverse event related to any patient's state of health. There is no patient directly associated with the survey sample. Internal biomerieux investigation was conducted. Investigational testing included the following. Vitek 2 yst id test kit confirmed the organism identification to candida albicans. Broth microdilution (bmd): fluconazole mic >64mg/l (r) and voriconazole mic>16mg/l (r). Vitek 2 ast-ys07 test kit (from sda media, customer lot and random lot): customer lot: fluconazole mic =16 mg/l (i) and voriconazole mic = 1 mg/l (s). Random lot: fluconazole mic =16 mg/l (i) and voriconazole mic = 1 mg/l (s). Vitek 2 ast-ys07 test kit (from cos media, customer lot): customer lot: fluconazole mic =<=1 mg/l (s) and voriconazole mic <= 0. 125 mg/l (s). Etest: fluconazole mic >= 256mg/l (r) and voriconazole mic >= 32 mg/l (r). It's possible that the environment of the vitek 2 card and shorter vitek 2 incubation time, compared to other methods (e. G. Bmd, etest), does not provide the optimal environment for this strain. The investigation concluded the (b)(6) survey organism is an atypical strain for vitek 2 testing and that the vitek 2 yeast (yst) identification (id) test kit is performing as intended.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1950204-2016-00115
MDR Report Key5921065
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-09-01
Date of Report2016-08-08
Date Mfgr Received2016-08-08
Device Manufacturer Date2015-09-26
Date Added to Maude2016-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. ELLEN WELTMER
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147317301
Manufacturer G1BIOMERIEUX, INC
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityST. LOUIS MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK? 2 AST-YS07 TEST KIT
Generic NameVITEK? 2 AST-YS07 TEST KIT
Product CodeNGZ
Date Received2016-09-01
Catalog Number414967
Lot Number287375612
Device Expiration Date2017-03-27
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMERIEUX, INC
Manufacturer Address595 ANGLUM ROAD ST. LOUIS MO 63042 US 63042


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.