THINPREP 5000 PROCESSOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-01 for THINPREP 5000 PROCESSOR manufactured by Hologic, Inc.

Event Text Entries

[54442117] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[54442118] A customer in (b)(6) reported their thinprep 5000 processor with autoloader printed the wrong number on the vial. All other vials in the same processing tray were checked and the numbers were found to be correct, no extra bar codes were found. The vial number 1620013697 was reloaded as a single sample on the t5 and was reprocessed correctly. No patient recall needed. Hologic's field service engineer (fse) confirmed but unable to reproduce the error. The barcode reader with linear base scanner is the most likely cause of the barcode misread. Performed the 2d barcode reader upgrade, processed samples to confirm operation. Instrument operational.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222780-2016-00222
MDR Report Key5922493
Date Received2016-09-01
Date of Report2016-09-01
Date Mfgr Received2016-09-01
Date Added to Maude2016-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSIDRA PIRACHAL
Manufacturer Street250 CAMPUS DRIVE
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTHINPREP 5000 PROCESSOR
Generic NameTHINPREP 5000 PROCESSOR
Product CodeMKQ
Date Received2016-09-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC, INC
Manufacturer Address250 CAMPUS DRIVE MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-09-01