PREP TRAY DRY SPONGE W/GLOVE 4461A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-01 for PREP TRAY DRY SPONGE W/GLOVE 4461A manufactured by Carefusion, Inc.

Event Text Entries

[54412941] (b)(4) a follow up emdr will be submitted upon completion of investigation or additional information received. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[54412942] Medwatch (b)(4) states, " surgeon preped the perineum with betadine stick sponges. The sponges came off in the vagina two times. They were retrieved by the surgeon. "
Patient Sequence No: 1, Text Type: D, B5


[58380126] (b)(4) neither the sample nor a photograph was received for evaluation. Therefore, the failure mode was unable to be confirmed. Master production records were reviewed for lot 914885 and no non-conformances were noted during the manufacturing of this lot. Our records indicate that the reviewed batch record passed all the in process inspection. During manufacturing of product assembly and packaging no special issues were noted and no ncr were issued to this lot number. Dhr of the sub-assembly was reviewed and no non-conformances were found. The most probable root cause was unable to be determined due to lack of sample. All internal manufacturing procedures were adhered to and no issues were found. The failure mode will be entered into the complaint management system and will be tracked / trended for any additional similar failure modes.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004932373-2016-00010
MDR Report Key5923300
Date Received2016-09-01
Date of Report2016-10-26
Date of Event2016-06-08
Date Mfgr Received2016-08-04
Date Added to Maude2016-09-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANNA WEHRHEIM
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer Phone8473628063
Manufacturer G1CAREFUSION MEXICALI MX PLANT
Manufacturer StreetCERRADA V
Manufacturer CityMEXICALI B.C;
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePREP TRAY DRY SPONGE W/GLOVE
Generic NameSKIN AND PREP TRAY
Product CodeOJU
Date Received2016-09-01
Model Number4461A
Lot Number0000914885
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address75 NORTH FAIRWAY DRIVE VERNON HILLS IL 60061 US 60061


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-01

© 2026 FDA.report