MERGE EYE STATION MERGE EYE STATION V11.1.38

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-09-06 for MERGE EYE STATION MERGE EYE STATION V11.1.38 manufactured by Merge Healthcare.

Event Text Entries

[54024960] Merge healthcare support helped the customer set up the new hard drive and the issue was resolved.
Patient Sequence No: 1, Text Type: N, H10


[54024961] Merge eye station is intended to be used in conjunction with existing ophthalmic fundus cameras to take images of the eye, perform fluorescein angiography, red free, color and icg still-image photography as well as video imaging. On (b)(6) 2016, merge healthcare was notified that a customer experienced a failed hard drive error message. Support sent out a replacement hard drive. On 08/09/2016, additional information was received from the customer that indicated a delay in examinations occurred as a result of the hard drive error. Specific information regarding the full impact could not be obtained. With merge eye station not performing as expected there is a potential for a delay in treatment or diagnosis that may result in harm to a patient. However, there is no indication of any patient harm as a result of this issue. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183926-2016-00720
MDR Report Key5931533
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-09-06
Date of Report2016-08-03
Date of Event2016-08-03
Date Mfgr Received2016-08-09
Date Added to Maude2016-09-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL DIEDRICK
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND, WI 53029
Manufacturer CountryUS
Manufacturer Postal53029
Manufacturer Phone2629123570
Manufacturer G1MERGE HEALTHCARE
Manufacturer Street900 WALNUT RIDGE DR.
Manufacturer CityHARTLAND, WI 53029
Manufacturer CountryUS
Manufacturer Postal Code53029
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMERGE EYE STATION
Generic NameCAMERA, OPHTHALMIC, AC-POWERED
Product CodeHKI
Date Received2016-09-06
Model NumberMERGE EYE STATION V11.1.38
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerMERGE HEALTHCARE
Manufacturer Address900 WALNUT RIDGE DRIVE HARTLAND, WI 53029 US 53029


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-06