VIVIDIMAGE T 26IN SURGICAL MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-08 for VIVIDIMAGE T 26IN SURGICAL MONITOR manufactured by Vts Medical Systems.

Event Text Entries

[54756090] A steris service technician inspected the monitor and assembly and found the power wire to be pinched in the spring arm which was subsequently making contact with another wire. The technician further inspected the assembly and found the ground wire in the ceiling from the monitor arm was not connected to the ceiling plate. The internal shield on the ground wire began to get hot and melted the wire coating, thus causing the reported smoke. Steris is unable to determine the cause of the pinched and disconnected wires. The user facility stated proper repairs would be completed on the assembly and returned to service. No further issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[54756091] The user facility reported that during cleaning of the monitor assembly the unit began to emit smoke. No patient involvement, no report of injury or procedural delays or cancellations.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000404456-2016-00007
MDR Report Key5937156
Date Received2016-09-08
Date of Report2016-09-08
Date of Event2016-08-09
Date Mfgr Received2016-08-09
Date Added to Maude2016-09-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1VTS MEDICAL SYSTEMS
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameVIVIDIMAGE T 26IN SURGICAL MONITOR
Generic NameSURGICAL MONITOR
Product CodeKQM
Date Received2016-09-08
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerVTS MEDICAL SYSTEMS
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.