GARDNER-WELLS TONG 66-1025

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-09 for GARDNER-WELLS TONG 66-1025 manufactured by Symmetry Surgical, Inc..

Event Text Entries

[54281352]
Patient Sequence No: 1, Text Type: N, H10


[54281353] The gardner-wells tongs were placed per manufacturer guidelines. The procedure was started and a shift in head position occurred. This alerted the staff that the tongs had actually slipped. The surgical team was able to control the head and maintain neck alignment while another set of tongs was obtained and reapplied and attached to traction. No further complications were noted. The stainless steel pinions appeared to have become dulled. This appeared to have affected the spring action in one of the pins that serves as a pressure indicator causing the tongs to slip on the skull. Manufacturer response for gardner-wells tong, symmetry surgical skull tong (per site reporter): they have no recommendations.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number5938789
MDR Report Key5938789
Date Received2016-09-09
Date of Report2016-08-22
Date of Event2016-07-25
Report Date2016-08-22
Date Reported to FDA2016-08-22
Date Reported to Mfgr2016-08-22
Date Added to Maude2016-09-09
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGARDNER-WELLS TONG
Generic NameTONG, SKULL FOR TRACTION
Product CodeHAX
Date Received2016-09-09
Model Number66-1025
Catalog Number66-1025
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSYMMETRY SURGICAL, INC.
Manufacturer Address3034 OWEN DRIVE ANTIOCH TN 37013 US 37013


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-09

© 2026 FDA.report