DIMENSION VISTA? SYSTEM K1021 SMN 10445159

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-09-09 for DIMENSION VISTA? SYSTEM K1021 SMN 10445159 manufactured by Siemens Healthcare Diagnostics Inc - Glasgow.

Event Text Entries

[54328964] The cause of the qc discrepant results is unknown. The account resolved the issue after instrument server 2 maintenance and recalibration. However, the pattern of low qc followed by high qc after recalibration is suggestive of vista bun field communication vc-16. 01. A. B. Ous which involved reagent lot 16055ab. Siemens healthcare diagnostics has determined that dimension vista? Blood urea nitrogen (bun) lots 15215ae, 15243bb, 15264ba, 15299bb, 15300ba, 15320bb, and 15341ac may exhibit inaccurate patient and/or quality control results. Only specific reagent cartridge wells are affected. If calibration is performed using an unaffected well and patient results are subsequently run using an affected well, bun results may be falsely depressed by up to approximately 50%. If calibration is performed using an affected well, bun results may be falsely elevated by up to approximately 64%. Siemens issued an urgent medical device correction (umdc) dated february 2016, communication vc-16. 01. B. Us to all u. S accounts or an urgent field safety notice vc-16-01. B. Ous, to all outside u. S. Accounts who had been shipped the impacted lots. Customers were directed to discard certain flexes with a specific lot number/cavity number combination. Siemens healthcare diagnostics confirmed additional complaints and has added additional blood urea nitrogen (bun k1021) lots 16004ab, 16034aa, 16033ac, 16048aa, 16055aa, 16055ab,16062aa, 16062ab as well as all future vista bun lots until further notice to the field action. An updated information communication umdc vc-16-01a. B. Us and ufsn vc-16-01a. B. Ous has been mailed in june 2016 to all accounts who have been shipped these lots. The communication advised customers to follow the additional instructions under the section "actions to be taken by the customer" to manage their bun testing. The account acknowledged receipt of the umdc communication. In a umdc update information - resolution communication sent to the customers dated july 22, 2016, siemens healthcare diagnostics communicated that investigation indicates loss of activity of the gldh enzyme reagent from specific mold cavities caused the issue. Not all flexes or well sets are impacted. Siemens has implemented additional pre-release testing for all vista bun flex reagent lots. The additional testing ensures acceptable lot performance including gldh activity. Beginning with dimension vista bun flex lot 16110ac (expiration date 2017-04-19), and including all subsequent lots that expire thereafter, customers will no longer be required to follow directions in umdc letters vc-16-01. B. Us and vc-16-01a. B. Us or the corresponding urgent field safety notices vc-16-01. B. Ous and vc-16-01a. B. Ous. The letter advised customers that once the laboratory has secured new inventory, they must discard older inventory. Starting with vista bun lot 16110ac, they may restore the normal use of vista bun flex reagent cartridges. It has been confirmed that the customer had received the siemens communications. The account resolved the issue after instrument server 2 maintenance and recalibration.
Patient Sequence No: 1, Text Type: N, H10


[54328965] The customer noted discrepant low bun qc results on the dimension vista flex(r) system. No patient samples were run while qc was out of laboratory range. The account recalibrated and qc recoveries were then elevated above the laboratory ranges. There is no indication that patient treatment was altered or prescribed on the basis of discrepant bun results. There was no report of adverse health consequences as a result of discrepant bun results.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2517506-2016-00324
MDR Report Key5939647
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-09-09
Date of Report2016-09-09
Date of Event2016-08-19
Date Mfgr Received2016-08-19
Device Manufacturer Date2016-02-25
Date Added to Maude2016-09-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer StreetGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE
Manufacturer CityNEWARK DE 19714
Manufacturer CountryUS
Manufacturer Postal Code19714
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-1622-2016
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION VISTA? SYSTEM
Generic NameDIMENSION VISTA? UREA NITROGEN FLEX? REAGENT CARTRIDGE,
Product CodeCDQ
Date Received2016-09-09
Catalog NumberK1021 SMN 10445159
Lot Number16055AB
Device Expiration Date2017-02-23
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Manufacturer AddressGLASGOW BUSINESS COMMUNITY 500 GBC DRIVE NEWARK DE 19714 US 19714


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.