OLYMPIC COOL-CAP SYSTEM 401351-501

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-09-09 for OLYMPIC COOL-CAP SYSTEM 401351-501 manufactured by Natus Medical Incorporated.

Event Text Entries

[54469420] Device has been returned to natus, and currently pending evaluation and investigation. When investigation is complete, and root cause determined, supplemental mdr will be submitted.
Patient Sequence No: 1, Text Type: N, H10


[54469421] Customer stated staff reported the cool cap display turned completely black while monitoring a baby. The staff rebooted and it has been working perfectly fine. No injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[66253177] Device was returned and evaluated. The first objective was to determine if the failure condition existed in the unit as-received, and if not, to see if it could be duplicated. The unit was turned on and it booted to the home screen normally. It was then run for 10 days, at which point a black screen failure occurred with the "operating system not found" message displayed. The alarm in the cooling module was beeping indicating that cooling had stopped. This event duplicated one of the failures reported by the hospital. The "operating system not found" error message is generated by the bios during the start-up sequence when the hard drive is actually missing, data is unusable, or it can't find the mbr. A compromised hard drive can result in the immediate reboot of the system. This has also been observed on lab units. This type of error is not uncommon in pc architecture-based systems, like the cool-cap system. The operating system is located on the hard drive, so if the hard drive is compromised, the system will fail to operate correctly. Because of the affinity towards a hard drive error in this unit, further tests were focused on hard drive integrity. The customer indicated that the unit worked properly after rebooting the system after the failure occurred. System logs indicate that it had been used several times between and after the reported failures. The unit ran 10 days at natus before the error occurred. Given these three scenarios, it can be concluded that the loss of the os is intermittent, and doesn't necessarily have a direct stimulus (such as an extreme environment, movement, etc). The black screen failure occurred twice at the hospital, and once at natus. All cases can be characterized as intermittent failures without a specific stimulus. Since the failure cannot be reproduced on demand, the exact root cause cannot be determined. However, the loss of the operating system is necessarily a result of compromised data from the hard drive. The most likely root cause is a failing hard drive or a failing ide cable, likely due to component age.
Patient Sequence No: 1, Text Type: N, H10


[66253178] Customer stated staff reported the cool cap display turned completely black while monitoring a baby. The staff rebooted and it has been working perfectly fine. No injury was reported. Patient related information was requested from the customer, but the customer has not responded to the inquiry.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3018859-2016-00013
MDR Report Key5941095
Report SourceUSER FACILITY
Date Received2016-09-09
Date of Report2016-01-11
Date Mfgr Received2016-01-11
Device Manufacturer Date2008-03-21
Date Added to Maude2016-09-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSVETLANA ROMANETS
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685170
Manufacturer G1NATUS MEDICAL INCORPORATED
Manufacturer Street5900 FIRST AVENUE SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal Code98108
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP SYSTEM
Generic NameCOOL CAP
Product CodeMXM
Date Received2016-09-09
Returned To Mfg2016-08-29
Model Number401351-501
OperatorNURSE
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer Address5900 FIRST AVENUE SOUTH SEATTLE WA 98108 US 98108


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-09-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.