MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-09-14 for PIN,PASSING,DRILL TIP 2.4X381MM,STRL 7208678 manufactured by Smith & Nephew, Inc..

Event Text Entries

[54635727] (b)(6). Evaluation was not possible, as the device was not returned. Review of the device history records was performed which confirmed no inconsistencies. There were no internal processing issues, which could have contributed to the nature of the complaint. A complaint history review has not identified additional complaints for this lot number on file. Due to these facts, we are unable to determine what may have caused the user to experience the reported incident. There are no indications that would suggest that the device did not meet product specifications upon release into distribution.
Patient Sequence No: 1, Text Type: N, H10


[54635728] It was reported that the tip of the passing pin broke off in the patient's knee while being passed through the femoral jig. The broken end was retrieved by the surgeon. No patient injury or other complications were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1219602-2016-00611
MDR Report Key5948393
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-09-14
Date of Report2014-07-30
Date of Event2014-07-21
Date Mfgr Received2014-07-30
Device Manufacturer Date2014-05-02
Date Added to Maude2016-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM GONZALES
Manufacturer Street150 MINUTEMAN ROAD
Manufacturer CityANDOVER 01810
Manufacturer CountryUS
Manufacturer Postal01810
Manufacturer Phone5123585706
Manufacturer G1SMITH & NEPHEW, INC.
Manufacturer Street130 FORBES BOULEVARD
Manufacturer CityMANSFIELD MA 02048
Manufacturer CountryUS
Manufacturer Postal Code02048
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePIN,PASSING,DRILL TIP 2.4X381MM,STRL
Generic NameGUIDE, DRILL, LIGAMENT
Product CodeLXI
Date Received2016-09-14
Catalog Number7208678
Lot Number50503116
Device Expiration Date2019-04-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW, INC.
Manufacturer Address130 FORBES BOULEVARD MANSFIELD MA 02048 US 02048


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-14