IVORY CLAMP SS 14A REG MOLAR 50057356

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a distributor report with the FDA on 2016-09-14 for IVORY CLAMP SS 14A REG MOLAR 50057356 manufactured by Heraeus Kulzer, Llc.

Event Text Entries

[54761056] This incident occurred in (b)(6). The office was using hu-friedy forceps, which can lead to hyperextension and excessive forces being placed on the bow of the ivory clamp. These forces and hyperextension can lead to the breakage of the clamp. Also noticed during the evaluation in the breakage area were a couple of small voids in the metal that could have contributed to this breakage. Due to the forceps used as described and the possible voids, this incident is impossible to conclude with 100% certainty which factor caused the malfunction.
Patient Sequence No: 1, Text Type: N, H10


[54761057] Clamp broke the first time in use. Clamp broke while on the forceps, getting ready to place into the mouth. Clamp broke outside of the patient's mouth. Office is using hu-friedy forceps (non heraeus kulzer forceps). A rubber dam was not being utilized for the procedure. Clamp was only ligated on one side.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1821514-2016-00027
MDR Report Key5951685
Report SourceDISTRIBUTOR
Date Received2016-09-14
Date of Report2016-09-12
Date of Event2016-08-01
Date Mfgr Received2016-08-16
Device Manufacturer Date2015-10-01
Date Added to Maude2016-09-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. AMBER BROWN
Manufacturer Street4315 LAFAYETTE BLVD.
Manufacturer CitySOUTH BEND IN 46614
Manufacturer CountryUS
Manufacturer Postal46614
Manufacturer Phone5742995411
Manufacturer G1HERAEUS KULZER, LLC
Manufacturer Street4315 LAFAYETTE BLVD.
Manufacturer CitySOUTH BEND IN 466142517
Manufacturer CountryUS
Manufacturer Postal Code466142517
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIVORY CLAMP SS 14A REG MOLAR
Generic NameRUBBER DAM CLAMP
Product CodeEEF
Date Received2016-09-14
Returned To Mfg2016-09-06
Catalog Number50057356
Lot NumberM5
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerHERAEUS KULZER, LLC
Manufacturer Address4315 LAFAYETTE BLVD. SOUTH BEND IN 466142517 US 466142517


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-09-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.