[54922003] Safety device that holds the nut in place that connects brakes to wheelchair did not keep nut in place despite following manufacturer's installation instructions. Position of bolt allows torque from brakes to loosen nut beyond the positioning of the safety cap that covers the nut. Wheel fell off wheelchair causing fall with head injury. All of these same brakes are removed from other wheelchairs. Upon removal of these same brakes from other wheelchairs, the nut was loose behind each of the orange safety caps. It appears that the orange safety cap covers approximately 1/4 inch or less of the bolt and the torque from braking results in a loosened nut with that doesn't engage the 1/4 inch of protection.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number5952916
MDR Report Key5952916
Date Received2016-09-07
Date of Report2016-09-09
Date of Event2016-09-04
Date Facility Aware2014-09-04
Report Date2016-09-09
Date Reported to FDA2016-09-09
Date Reported to Mfgr2016-09-09
Date Added to Maude2016-09-15
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameTHE AFTERMARKET GROUP
Generic NameSECURE AUTOMATIC WHEEL LOCK KIT
Product CodeKNO
Date Received2016-09-07
OperatorLAY USER/PATIENT
Device AvailabilityY
Device Age30 DA
Device Eval'ed by MfgrI
Device Sequence No1
Device Event Key0
ManufacturerUNITED PLASTIC MOLDERS, INC.
Manufacturer AddressJACKSON MS US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Life Threatening 2016-09-07