COBAS E-ELECSYS ANTI-CCP 05031656190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-09-15 for COBAS E-ELECSYS ANTI-CCP 05031656190 manufactured by Roche Diagnostics.

Event Text Entries

[54894089] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[54894090] The customer received questionable elecsys anti-ccp immunoassay results for one patient sample from a cobas e602 analyzer (unknown serial number) and another unknown roche analyzer. The initial result was 197 u/ml (positive). The sample was tested on another roche system and the result was positive. No specific data was provided. The result on an abbott architect on (b)(6) 2016 was 6. 0 u/ml (negative). The result with elisa (euroimmun) on 07/08/2016 had been <5 u/ml (negative). The result with thermofisher (elia) was negative. No specific data or date of testing was provided. Information concerning if the erroneous result was reported the patient and/or medical personnel treating the patient was requested, but it was unknown. The patient was not adversely affected. As no sample material was available for further investigation, a specific root cause could not be determined.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1823260-2016-01386
MDR Report Key5953335
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-09-15
Date of Report2016-09-15
Date of Event2016-07-12
Date Mfgr Received2016-08-25
Date Added to Maude2016-09-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS E-ELECSYS ANTI-CCP
Generic NameANTIBODIES, ANTI-CYCLIC CITRULLINATED PEPTIDE (CCP)
Product CodeNHX
Date Received2016-09-15
Model NumberNA
Catalog Number05031656190
Lot Number13592000
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.