MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2004-06-17 for PALL EBDS NA manufactured by Pall Medsep.
[389185]
During a product qualification at a user facility, prior to clinical implementation, unexpected results were obtained. Using purposely-contaminated samples as a positive control, the instrument initially correctly determined that the sample was contaminated, but on subsequent determinations, did not indicate such pouches to be contaminated. A second instrument was tested, and consistently gave the expected determination indication contamination.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2013342-2004-00005 |
MDR Report Key | 595557 |
Report Source | 05,06,07 |
Date Received | 2004-06-17 |
Date of Report | 2004-04-27 |
Date of Event | 2004-04-22 |
Date Facility Aware | 2004-04-22 |
Date Reported to Mfgr | 2004-04-27 |
Date Mfgr Received | 2004-04-27 |
Date Added to Maude | 2005-04-27 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 0 |
Manufacturer Contact | LEONARD BERMAN, PH.D |
Manufacturer Street | 2200 NORTHERN BLVD |
Manufacturer City | EAST HILLS NY 11548 |
Manufacturer Country | US |
Manufacturer Postal | 11548 |
Manufacturer Phone | 5164845400 |
Manufacturer G1 | PALL MEDSEP |
Manufacturer Street | 1630 INDUSTRIAL PARK ST. |
Manufacturer City | COVINA CA 91722 |
Manufacturer Country | US |
Manufacturer Postal Code | 91722 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | PALL EBDS |
Generic Name | BACTERIA DETECTION SYSTEM SAMPLE SET |
Product Code | MZC |
Date Received | 2004-06-17 |
Model Number | EBDS |
Catalog Number | NA |
Lot Number | SET L/N 0400568 |
ID Number | * |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Eval'ed by Mfgr | Y |
Implant Flag | N |
Date Removed | * |
Device Sequence No | 1 |
Device Event Key | 585383 |
Manufacturer | PALL MEDSEP |
Manufacturer Address | * COVINA CA * US |
Baseline Brand Name | PALL EBDS |
Baseline Generic Name | BACTERIA DETECTION SYSTEM SAMPLE SET |
Baseline Model No | EBDS |
Baseline Catalog No | NA |
Baseline ID | * |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2004-06-17 |