PALL EBDS NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2004-06-17 for PALL EBDS NA manufactured by Pall Medsep.

Event Text Entries

[389185] During a product qualification at a user facility, prior to clinical implementation, unexpected results were obtained. Using purposely-contaminated samples as a positive control, the instrument initially correctly determined that the sample was contaminated, but on subsequent determinations, did not indicate such pouches to be contaminated. A second instrument was tested, and consistently gave the expected determination indication contamination.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2013342-2004-00005
MDR Report Key595557
Report Source05,06,07
Date Received2004-06-17
Date of Report2004-04-27
Date of Event2004-04-22
Date Facility Aware2004-04-22
Date Reported to Mfgr2004-04-27
Date Mfgr Received2004-04-27
Date Added to Maude2005-04-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLEONARD BERMAN, PH.D
Manufacturer Street2200 NORTHERN BLVD
Manufacturer CityEAST HILLS NY 11548
Manufacturer CountryUS
Manufacturer Postal11548
Manufacturer Phone5164845400
Manufacturer G1PALL MEDSEP
Manufacturer Street1630 INDUSTRIAL PARK ST.
Manufacturer CityCOVINA CA 91722
Manufacturer CountryUS
Manufacturer Postal Code91722
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePALL EBDS
Generic NameBACTERIA DETECTION SYSTEM SAMPLE SET
Product CodeMZC
Date Received2004-06-17
Model NumberEBDS
Catalog NumberNA
Lot NumberSET L/N 0400568
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key585383
ManufacturerPALL MEDSEP
Manufacturer Address* COVINA CA * US
Baseline Brand NamePALL EBDS
Baseline Generic NameBACTERIA DETECTION SYSTEM SAMPLE SET
Baseline Model NoEBDS
Baseline Catalog NoNA
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2004-06-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.