STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 229 777400-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-16 for STATIC AND DYNAMIC COMPLIANCE VMAX ENCORE 229 777400-101 manufactured by Carefusion, Inc.

Event Text Entries

[55259963] (b)(4). Carefusion field service representative (fsr) was dispatched to perform an onsite evaluation. The fsr performed a firmware upgrade on the treadmill to 0. 4f. The fsr then performed an operation check and ran a custom protocol in order to verify the functionality of the treadmill. The testing was run five times and the reported issue was resolved. The treadmill is meeting all manufacture specifications no device or parts will be returned at this time.
Patient Sequence No: 1, Text Type: N, H10


[55259964] The customer reported that while using the static and dynamic compliance, the track master treadmill tmx428cp stopped in the middle of a test. The customer reported seeing a? Tether error? Message. The reported issue occurred during preparation for use and there was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2021710-2016-04453
MDR Report Key5956492
Date Received2016-09-16
Date of Report2016-08-25
Date of Event2016-08-25
Date Mfgr Received2016-08-25
Device Manufacturer Date2015-09-01
Date Added to Maude2016-09-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRESPIRATORY THERAPIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJILL RITTORNO
Manufacturer Street22745 SAVI RANCH PARKWAY
Manufacturer CityYORBA LINDA CA 92887
Manufacturer CountryUS
Manufacturer Postal92887
Manufacturer Phone7149227830
Manufacturer G1CAREFUSION, INC
Manufacturer Street1100 BIRD CENTER DRIVE
Manufacturer CityPALM SPRINGS CA 92262
Manufacturer CountryUS
Manufacturer Postal Code92262
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameSTATIC AND DYNAMIC COMPLIANCE
Generic NameCALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Product CodeBTY
Date Received2016-09-16
Model NumberVMAX ENCORE 229
Catalog Number777400-101
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION, INC
Manufacturer Address22745 SAVI RANCH PARKWAY YORBA LINDA CA 92887 US 92887


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.