COULTER TQ-PREP WORKSTATION N/A 6605429

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-09-16 for COULTER TQ-PREP WORKSTATION N/A 6605429 manufactured by Beckman Coulter.

Event Text Entries

[54995900] The customer reported a bad scatter pattern when samples were prepared on their tq prep instrument. The fse (field service engineer) observed a failure of the sensors. No error message alerted the customer that the reagent bottles were empty since the sensors had failed. The fse replaced the reagent a and reagent b level sense assemblies to resolve the issue. Repairs were verified per service procedures. The beckman coulter internal identifier for this event is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[54995902] The customer reported a bad scatter pattern when samples were prepared on their coulter tq-prep workstation. Erroneous results were not generated and there was no change or effect to patient treatment in connection with this event. Customer stopped using the instrument as soon as the problem was detected.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1061932-2016-00815
MDR Report Key5957849
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-09-16
Date of Report2016-08-23
Date of Event2016-08-23
Date Mfgr Received2016-08-23
Device Manufacturer Date2006-08-01
Date Added to Maude2016-09-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MIRANDA HOLLAND
Manufacturer Street11800 SW 147TH AVENUE, 32-S08
Manufacturer CityMIAMI FL 331962031
Manufacturer CountryUS
Manufacturer Postal331962031
Manufacturer Phone3053802031
Manufacturer G1BECKMAN COULTER
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCOULTER TQ-PREP WORKSTATION
Generic NameAUTOMATED PIPETTING, DILUTING AND SPECIMEN PROCESSING WORKSTATIONS FOR FLOW CYTO
Product CodePER
Date Received2016-09-16
Model NumberN/A
Catalog Number6605429
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.