MERGE EYE STATION MERGE EYE STATION V11.5

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-09-18 for MERGE EYE STATION MERGE EYE STATION V11.5 manufactured by Merge Healthcare.

Event Text Entries

[55002311] Merge healthcare is currently investigating this customer's allegation to determine if corrections or corrective actions are required.
Patient Sequence No: 1, Text Type: N, H10


[55002312] Merge eye station is intended to be used in conjunction with existing ophthalmic fundus cameras to take images of the eye, perform fluorescein angiography, red free, color and icg still-image photography as well as video imaging. (b)(6) 2016, merge healthcare received information indicating a camera flash was not working. Support shipped a sync box and cables. After additional troubleshooting, it was determined the issue was occurring from a camera setting. The issue was resolved. On (b)(6) 2016, the customer reported the issue led to a delay in testing because of daily patient volume and the inability to image in a timely manner. There was no indication of patient harm; however, this issue is being reported due to the potential for harm related to a delay in testing. Reference complaint number (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183926-2016-00734
MDR Report Key5959188
Report SourceUSER FACILITY
Date Received2016-09-18
Date of Report2016-08-18
Date of Event2016-08-18
Date Mfgr Received2016-08-22
Date Added to Maude2016-09-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MIKE DIEDRICK
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND WI 53029
Manufacturer CountryUS
Manufacturer Postal53029
Manufacturer G1MERGE HEALTHCARE
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND WI 53029
Manufacturer CountryUS
Manufacturer Postal Code53029
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMERGE EYE STATION
Generic NameCAMERA, OPHTHALMIC, AC-POWERED
Product CodeHKI
Date Received2016-09-18
Model NumberMERGE EYE STATION V11.5
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMERGE HEALTHCARE
Manufacturer Address900 WALNUT RIDGE DRIVE HARTLAND, WI 53029 US 53029


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-18

© 2026 FDA.report