ULTRA DRIVE TOOLS 9.5MM SHORT OSTEOTOME N/A 423880

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-09-20 for ULTRA DRIVE TOOLS 9.5MM SHORT OSTEOTOME N/A 423880 manufactured by Biomet Sports Medicine.

Event Text Entries

[55142519] (b)(4). The product identification necessary to review manufacturing history was not provided. Visual inspection of the returned device found distal end damage and gouges. The distal end is charred, suggesting possible improper techniques and/or a lack of required irrigation during use. These situations may contribute to over-heating and fracture. Fracture surfaces of the osteotome suggest multiple origin fatigue fractures. A conclusive root cause of the event could not be determined with available information. The following could not be completed with the limited information provided. Date of event - ni; no information was provided about the fracture of the osteotome; however, it may have occurred during another event. This event is reported on 1825034-2016-03461. Manufacture date? Ni.
Patient Sequence No: 1, Text Type: N, H10


[55142520] Osteotome was found to be fractured. It is unknown if the device fractured during a procedure. No further information has been provided.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001825034-2016-03721
MDR Report Key5963431
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-09-20
Date of Report2016-08-23
Date Mfgr Received2016-08-23
Date Added to Maude2016-09-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MEGAN HAAS
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5743726700
Manufacturer G1BIOMET ORTHOPEDICS
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameULTRA DRIVE TOOLS 9.5MM SHORT OSTEOTOME
Generic NameDEVICE, NERVE CONDUCTION VELOCITY MEASUREMENT
Product CodeJXE
Date Received2016-09-20
Returned To Mfg2016-08-23
Model NumberN/A
Catalog Number423880
Lot NumberNI
ID NumberN/A
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBIOMET SPORTS MEDICINE
Manufacturer Address56 E. BELL DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-20

© 2026 FDA.report