PROCOL BIOLOGIC VASCULAR GRAFT, 6MM X 40 CM HJL016-40-N

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-22 for PROCOL BIOLOGIC VASCULAR GRAFT, 6MM X 40 CM HJL016-40-N manufactured by Lemaitre Vascular, Inc..

Event Text Entries

[55720195] The device was received for evaluation but since it had been prepped for use and exposed to blood/preservatives we were unable to reach any conclusions from our investigation. The device history record for this lot was reviewed. The lot passed all inspection criteria including: visual inspection, external diameter, and leak testing at simulated physiological pressure. This is a bioprosthetic device and variation along the graft is normal. Our ifu does state that a "small amount of oozing is acceptable". At this time we are unable to determine the cause of the incident but believe it was an isolated event.
Patient Sequence No: 1, Text Type: N, H10


[55720196] During implantation of the graft, the surgeon pressurized it and noticed that a one inch section near the end appeared thin, lighter in color and started to slightly leak. The doctor removed the graft and used a different graft in stock without further incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1220948-2016-00021
MDR Report Key5970207
Date Received2016-09-22
Date of Report2016-09-22
Date of Event2016-08-23
Date Mfgr Received2016-08-24
Device Manufacturer Date2015-07-24
Date Added to Maude2016-09-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. PRAGYA THIKEY
Manufacturer Street63 SECOND AVE
Manufacturer CityBURLINGTON MA 01803
Manufacturer CountryUS
Manufacturer Postal01803
Manufacturer Phone7812212266
Manufacturer G1HANCOCK JAFFE LABS
Manufacturer Street70 DOPPLER
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NamePROCOL BIOLOGIC VASCULAR GRAFT, 6MM X 40 CM
Generic NameTISSUE GRAFT
Product CodeLXA
Date Received2016-09-22
Returned To Mfg2016-08-28
Model NumberHJL016-40-N
Device Expiration Date2019-12-30
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerLEMAITRE VASCULAR, INC.
Manufacturer Address63 SECOND AVE BURLINGTON MA 01803 US 01803


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-09-22

© 2026 FDA.report