NAVIOS FLOW CYTOMETER SYSTEM A52103

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-09-23 for NAVIOS FLOW CYTOMETER SYSTEM A52103 manufactured by Beckman Coulter Ireland.

Event Text Entries

[55490453] A field service engineer (fse) evaluated the instrument. The fse found that tubing from the valve connector was leaking. The fse replaced the tubing and the sheath sensor, which repaired the leak. The repairs were verified by the fse. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[55490454] The customer reported a leak from the navios flow cytometer. The volume of the leak was about 500 ml and was not contained within the instrument. The customer was wearing personal protective equipment (ppe) consisting of laboratory coat and gloves at the time of the event. There was no report of injury or biohazard exposure to open wounds or mucous membranes. Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
Patient Sequence No: 1, Text Type: D, B5


[58369858] The device date of manufacture was changed from 12/01/2009 to 07/01/2012.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2016-00841
MDR Report Key5974011
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-09-23
Date of Report2016-08-31
Date of Event2016-08-30
Date Mfgr Received2016-08-31
Device Manufacturer Date2012-07-01
Date Added to Maude2016-09-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MIRANDA HOLLAND
Manufacturer Street11800 SW 147TH AVE 32-L23
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal33196
Manufacturer Phone3053802031
Manufacturer G1BECKMAN COULTER
Manufacturer Street11800 SW 147TH AVE
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameNAVIOS FLOW CYTOMETER SYSTEM
Generic NameFLOW CYTOMETRIC REAGENTS AND ACCESSORIES
Product CodeOYE
Date Received2016-09-23
Model NumberNA
Catalog NumberA52103
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER IRELAND
Manufacturer Address250 SOUTH KRAEMER BLVD. LISMEEHAN O'CALLAGHAN'S MILLS COUNTY CLARE, EI


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.