VIVIDIMAGE T 26IN SURGICAL MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-23 for VIVIDIMAGE T 26IN SURGICAL MONITOR manufactured by Steris Corporation - Vts Medical Systems.

Event Text Entries

[55887053] A karl storz camera was being utilized during the time of the reported event. The image from the camera was being displayed on the steris vivid image monitor during the patient procedure. A steris service technician inspected the monitor and found a compatibility issue between the karl storz camera and steris vivid image monitor. Steris is working with the user facility to resolve the reported event. Investigation of this event is currently in process.
Patient Sequence No: 1, Text Type: N, H10


[55887054] The user facility reported that during a patient procedure the image displayed on the monitor froze. A procedural delay occurred due to the reported event. No report of injury.
Patient Sequence No: 1, Text Type: D, B5


[62413898] A steris service technician inspected the steris vividimage monitor and system and made repairs so the monitor properly displayed the image from the (b)(6) camera. The steris technician tested the system and (b)(6) camera and confirmed them to be operating properly. No additional compatibility issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1000404456-2016-00008
MDR Report Key5974050
Date Received2016-09-23
Date of Report2016-09-23
Date of Event2016-08-24
Date Mfgr Received2016-08-24
Date Added to Maude2016-09-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CORPORATION - VTS MEDICAL SYSTEMS
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVIVIDIMAGE T 26IN SURGICAL MONITOR
Generic NameSURGICAL MONITOR
Product CodeKQM
Date Received2016-09-23
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - VTS MEDICAL SYSTEMS
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.