CHOL2 CHOLESTEROL GEN.2 05168538190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-09-27 for CHOL2 CHOLESTEROL GEN.2 05168538190 manufactured by Roche Diagnostics.

Event Text Entries

[55722540] This event occurred in (b)(6). (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[55722541] The customer received questionable chol2 cholesterol gen. 2 results for an unknown number of patient samples from cobas c701 serial number (b)(4). The customer had performed qc which was in range, so they began testing patient samples. Two hours later, the customer noticed high results (about 2 g/l higher) compared to the patient samples performed in the morning. The customer repeated qc which was high, replaced the reagent pack, recalibrated, and performed qc which was acceptable. The samples that had high results were repeated with the new reagent pack. Data was provided for five patient samples as examples. Patient 1: initial result 4. 19 g/l, repeat result 2. 27 g/l. Patient 2: initial result 3. 69 g/l, repeat result 1. 73 g/l. Patient 3: initial result 4. 19 g/l, repeat result 2. 22 g/l. Patient 4: initial result 3. 84 g/l, repeat result 1. 85 g/l. Patient 5: initial result 4. 92 g/l, repeat result 2. 31 g/l. Information concerning if any erroneous results was reported outside the laboratory was requested, but it was not unknown. The patients were not adversely affected. At the end of the day, the customer repeated qc on the suspect reagent pack and the results were lower than those received earlier in the day. It was noted when the high results were discovered, half of the reagent pack had been consumed. The suspected reagent was inspected and it was found that the color was different between compartment b and c. Compartment c had a darker color than compartment b. It was suspected that oxidation of compartment c had occurred.
Patient Sequence No: 1, Text Type: D, B5


[63494551] The cholesterol reagent cassette has the same reagent in both compartments b and c. Tests are first pipetted out of compartment b (around 1500 tests) and then the instrument starts pipetting out of compartment c. The investigation suspected that while opening the cassette, something contaminated compartment c leading to an oxidation of the reagent as only the reagent in compartment c was affected. A blank calibration and qc done with compartment b was acceptable comparable to the previous calibrations. When the instrument switched to compartment c, the qc and patient samples recovered too high due to the oxidation. The following full calibration was then probably pipetted out of compartment c and the high standard 1 absorbance was seen. As no other reagent cassettes of this lot showed this behavior, it was assumed this was a single incidence and did not occur during production. No other similar complaints with this or any other lot of the reagent have been received.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2016-01472
MDR Report Key5979855
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-09-27
Date of Report2016-12-20
Date of Event2016-09-15
Date Mfgr Received2016-09-15
Date Added to Maude2016-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCHOL2 CHOLESTEROL GEN.2
Generic NameENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL
Product CodeCHH
Date Received2016-09-27
Model NumberNA
Catalog Number05168538190
Lot Number145482
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 462500457 US 462500457


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.