SORIN S5 BUBBLE DETECTOR FOR 3/8 TUBING 23-07-50

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-09-27 for SORIN S5 BUBBLE DETECTOR FOR 3/8 TUBING 23-07-50 manufactured by Sorin Group Deutschland.

Event Text Entries

[55785464] Sorin group (b)(4) manufactures the sorin s5 bubble detector. The incident occurred in (b)(6). This medwatch report is being filed on behalf of sorin group (b)(4). Sorin group (b)(4) received a report that the sorin s5 bubble detector was found to have deflated cushions. This issue was discovered by a sorin group field service representative during service. There was no patient involvement. The service representative replaced the bubble sensor and performed a functional check and trial run. No further issues were discovered. The investigation is ongoing. A follow-up report will be sent when the investigation is complete.
Patient Sequence No: 1, Text Type: N, H10


[55785465] Sorin group (b)(4) received a report that the sorin s5 bubble detector was found to have deflated cushions. This issue was discovered by a sorin group field service representative during service. There was no patient involvement.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611109-2016-00678
MDR Report Key5982188
Report SourceHEALTH PROFESSIONAL
Date Received2016-09-27
Date of Report2015-10-15
Date of Event2015-10-15
Date Mfgr Received2015-10-15
Device Manufacturer Date2010-01-21
Date Added to Maude2016-09-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactCARRIE WOOD
Manufacturer Street14401 W 65TH.WAY
Manufacturer CityARVADA CO 80004
Manufacturer CountryUS
Manufacturer Postal80004
Manufacturer Phone3034676461
Manufacturer G1SORIN GROUP DEUTSCHLAND
Manufacturer StreetLINDBERGHSTR. 25
Manufacturer CityMUNICH, 80939
Manufacturer CountryGM
Manufacturer Postal Code80939
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSORIN S5 BUBBLE DETECTOR FOR 3/8 TUBING
Generic NameMONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS
Product CodeKRL
Date Received2016-09-27
Model Number23-07-50
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSORIN GROUP DEUTSCHLAND
Manufacturer AddressLINDBERGHSTR. 25 MUNICH, 80939 GM 80939


Patients

Patient NumberTreatmentOutcomeDate
10 2016-09-27

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.