HOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE WITH INSPECTION PORT DFH-0008

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-28 for HOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE WITH INSPECTION PORT DFH-0008 manufactured by Microsurgical Technology, Inc..

Event Text Entries

[56357466] The device evaluation showed that the blade was broken about midway on the tine. Broken blade was not returned. Instrument was returned dirty, with dried residue on the exterior and interior. Small rust spots visible, and remaining blade appears rusty with a corrosion pit on the surface. Fractured surface of the broken tine has a darkened spot, which may correspond with prior corrosion pits. Cannula is slightly bent at base. Device failure is due to corrosion as a result of poor maintenance.
Patient Sequence No: 1, Text Type: N, H10


[56357467] The surgeon reported that the blade "fell off" in the eye during surgery. The blade was removed and there was no known harm to the patient.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3019924-2016-00017
MDR Report Key5984254
Date Received2016-09-28
Date of Report2016-09-26
Date of Event2016-08-25
Date Mfgr Received2016-09-07
Device Manufacturer Date2013-03-06
Date Added to Maude2016-09-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. JOHN RAY
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4255560544
Manufacturer G1MICROSURGICAL TECHNOLOGY, INC.
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHOFFMAN/AHMED HORIZONTAL MICRO-SCISSORS, CURVED, 23 GAUGE WITH INSPECTION PORT
Generic NameOPHTHALMIC SCISSORS
Product CodeHNF
Date Received2016-09-28
Returned To Mfg2016-09-16
Model NumberDFH-0008
Catalog NumberDFH-0008
Lot Number051735
OperatorPHYSICIAN
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY, INC.
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-09-28

© 2026 FDA.report