EASY GO VAC PM65

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-09-29 for EASY GO VAC PM65 manufactured by Precision Medical Inc..

Event Text Entries

[56544743] (b)(6). Subject: easy go aspirator serial # (b)(4). Med watch: (b)(4). Dear (b)(6), the pm65 serial number (b)(4), found with a broken power jack. The history on the device is that it was manufactured in october of 2004, and precision medical shows no record of the device being returned for repair. The device is 12 years old, or twice the recorded life of the product. The likely cause of the complaint, is that the male connector that mates with the device was forced into the connector, braking the female internal power connector. This happens as a result of dropping the device, or the age of the device caused the socket to fail. I would recommend that all pm65 aspirators be checked for this reason, to be sure that the internal battery is being charged. The life of the pm65 aspirator is 5 years. The warranty of the pm65 is 2 years. The life cycle of a lead acid battery is approximately 2 years. I would also recommend that pm65 devices, twice the average life span, be considered for replacement. Any other questions please feel free to contact me. Regards, (b)(4), quality manager, precision medical inc. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[56544744] Event description: audit completed by biomed. Approximately 2 months ago -2 other precision pm65 suction machines not functioning correctly, with same issue, found in 2 different units. An audit at sister hospital # 1 found 2 portable suction machines not functioning correctly. No defective units found at sister hospital # 2 as of 2 weeks after initial audit. Power adapter connector was found to be broken internally preventing charging of the battery or use when plugged into a ac outlet. This indication would not be observable from the outside of the device. The connector will be replaced. During further investigation of our fleet of portable suctions: two additional devices were found with the same issue; these two will be repaired. The two suction mentioned at sister hospital #1 were a different issue; they had the wrong voltage charger.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523148-2016-00002
MDR Report Key5986779
Date Received2016-09-29
Date of Report2016-06-28
Date of Event2016-04-17
Report Date2016-06-28
Date Reported to FDA2016-06-28
Date Reported to Mfgr2016-08-12
Date Mfgr Received2016-08-12
Device Manufacturer Date2004-10-04
Date Added to Maude2016-09-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR JAMES PARKER
Manufacturer Street300 HELD DRIVE
Manufacturer CityNORTHAMPTON PA 18067
Manufacturer CountryUS
Manufacturer Postal18067
Manufacturer Phone6102626090
Manufacturer G1PRECISION MEDICAL INC
Manufacturer Street300 HELD DRIVE
Manufacturer CityNORTHAMPTON PA 18067
Manufacturer CountryUS
Manufacturer Postal Code18067
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameEASY GO VAC
Generic NamePUMP.PORTABLE, ASPIRATION (MANUAL OR POWERED)
Product CodeBTA
Date Received2016-09-29
Model NumberPM65
Catalog NumberPM65
OperatorBIOMEDICAL ENGINEER
Device Availability*
Device Age12 YR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPRECISION MEDICAL INC.
Manufacturer Address300 HELD DRIVE NORTHAMPTON PA 18067 US 18067


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-09-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.