OPMI PICO ON S100 FLOOR STAND 000000-1245-414

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-09-29 for OPMI PICO ON S100 FLOOR STAND 000000-1245-414 manufactured by Carl Zeiss Meditec Ag (oberkochen).

Event Text Entries

[55934105] A field service engineer (fse) performed an on-site inspection and found that the two screws of the mounting plate securing the xenon power supply illumination unit to the s100 floor stand were broken. The fse repaired the holding mecahnism and put the microscope back into service.
Patient Sequence No: 1, Text Type: N, H10


[55934106] The health care professional reported the following: when switching off the opmi pico microscope on an s100 floor stand after finishing the oral procedure, the nurse noticed that the lamp housing was loose and only hanging on one screw. Then, the lamp housing came completely loose and the nurse was able to catch the lamp housing. The part did not fall on the floor nor hit anyone. The procedure had already been completed and no patient was involved. No one was injured due to this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9615010-2016-00014
MDR Report Key5987231
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-09-29
Date of Report2016-09-29
Date of Event2016-08-31
Date Mfgr Received2016-09-02
Device Manufacturer Date2013-12-16
Date Added to Maude2016-09-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. WILLIAM GUSTAFSON
Manufacturer Street5160 HACIENDA DRIVE
Manufacturer CityDUBLIN CA 94568
Manufacturer CountryUS
Manufacturer Postal94568
Manufacturer Phone9255574689
Manufacturer G1CARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer StreetRUDOLF-EBER-STRASSE 11
Manufacturer CityOBERKOCHEN, BADEN-WUERTTEMBERG 73447
Manufacturer CountryGM
Manufacturer Postal Code73447
Single Use3
Remedial ActionRP
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameOPMI PICO ON S100 FLOOR STAND
Generic NameOPMI PICO MICROSCOPE
Product CodeFSO
Date Received2016-09-29
Model NumberNA
Catalog Number000000-1245-414
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCARL ZEISS MEDITEC AG (OBERKOCHEN)
Manufacturer AddressRUDOLF-EBER-STRASSE 11 OBERKOCHEN, BADEN-WUERTTEMBERG 73447 GM 73447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-09-29