SURGIGUIDE 37503

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional report with the FDA on 2016-10-06 for SURGIGUIDE 37503 manufactured by Dentsply Implants N.v..

Event Text Entries

[56590876] Since this event resulted in medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the malfunction would be likely to cause or contribute to a serious injury should it recur. As such, this event meets the definition of a reportable event per 21 cfr part 803. Evaluation found there is scatter in the images and no clear outline of the mucosa information. There is nothing wrong with the fit on the plaster model or with the design in the planning file.
Patient Sequence No: 1, Text Type: N, H10


[56590877] It was reported that a surgiguide was used to create the osteotomy for three dental implants (tooth position 5, 6 and 7). After the first drilling step the surgical guide was removed to complete the additional drilling steps free handed according to the pilot guide drilling protocol. After placing the implants free handed, the customer discovered that the three implants were positioned too buccal. A second surgery is needed to correct the positioning of the implants.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3007362683-2016-00012
MDR Report Key6003092
Report SourceHEALTH PROFESSIONAL
Date Received2016-10-06
Date of Report2016-08-29
Date of Event2016-02-29
Date Mfgr Received2016-08-29
Date Added to Maude2016-10-06
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS. HELEN LEWIS
Manufacturer Street221 W. PHILADELPHIA ST. SUITE 60W
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178494229
Manufacturer G1DENTSPLY IMPLANTS N.V.
Manufacturer StreetRESEARCH CAMPUS 10
Manufacturer CityHASSELT LIMBURG, B-3500
Manufacturer CountryBE
Manufacturer Postal CodeB-3500
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameSURGIGUIDE
Generic NameSYSTEM, IMAGE PROCESSING, RADIOLOGICAL
Product CodeLLZ
Date Received2016-10-06
Returned To Mfg2016-09-12
Model NumberNA
Catalog Number37503
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Sequence No1
Device Event Key0
ManufacturerDENTSPLY IMPLANTS N.V.
Manufacturer AddressRESEARCH CAMPUS 10 HASSELT LIMBURG, B-3500 BE B-3500

Device Sequence Number: 2

Product CodeDZE
Date Received2016-10-06
Device Sequence No2
Device Event Key0

Device Sequence Number: 3

Product CodeEBG
Date Received2016-10-06
Device Sequence No3
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-10-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.