INTEGRATOR MALLET 015-99603

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-10-07 for INTEGRATOR MALLET 015-99603 manufactured by Integra York, Pa Inc..

Event Text Entries

[56931847] Integra has completed their internal investigation on (b)(6) 2016. The investigation included: evaluation of actual device, review of device history records, review of complaints history. The returned integrator showing wears, repair markings and a broken center screw. The integrator was manually tested and the center screw is broken and does not function as designed. The complaint is confirmed; unsuccessful repair nonconforming product report / nonconforming material report history: none. Complaints metrics tracked, trended, and reviewed monthly with management. There was one intergrator mallet repaired (b)(6) 2016 returned used/processed showing wears, repair markings and a broken center screw. The complaint report has been confirmed as a repair issue; we have taken measures to address this issue.
Patient Sequence No: 1, Text Type: N, H10


[56931848] It was reported that the integrator mallet is not cocking after being repaired for the same failure six days earlier. Product was in contact with the patient however, no patient injury was reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2016-00165
MDR Report Key6010883
Report SourceUSER FACILITY
Date Received2016-10-07
Date of Report2016-09-14
Date Mfgr Received2016-09-14
Date Added to Maude2016-10-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SONIA IRIZARRY
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA YORK, PA INC.
Manufacturer Street589 DAVIES DRIVE
Manufacturer CityYORK PA 17402
Manufacturer CountryUS
Manufacturer Postal Code17402
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameINTEGRATOR MALLET
Generic NameM54 - ORAL SURGERY
Product CodeLXM
Date Received2016-10-07
Returned To Mfg2016-09-21
Catalog Number015-99603
Lot Number1606R (1789)
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA YORK, PA INC.
Manufacturer Address589 DAVIES DRIVE 589 DAVIES DRIVE YORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.