PROVOX VEGA XTRASEAL 7782-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-10-10 for PROVOX VEGA XTRASEAL 7782-01 manufactured by Atos Medical Ab.

Event Text Entries

[56920528] This event is reportable because in the event of reoccurrence, the device could fall into the trachea with risk for clinical complications. Investigation: the voice prosthesis was not returned, an investigation of the product was therefore not possible. Based on the information given, there is no good reason to believe there has been something wrong with the product. However; it cannot be excluded that the products may have been damaged during the insertion. Discussion: a dislodged prosthesis is most often related to the quality of the fistula. E. G. A fistula that is created with a scalpel can become keyhole shaped. A weak or miss-shaped fistula gives less retention force towards the voice prosthesis t-flange and e-flange. This means that the voice prosthesis will be more prone to dislodge. Miss-shaped fistulas that need short prostheses have a higher risk of suffering this problem. The design of the flanges is made so that retention force on the esophagus flange is significant larger than for the tracheal flange. This has been done to prevent the prosthesis ending up in the trachea. Conclusion/action: there is no reason to believe that the dislodged prosthesis is due to product fault. The ifu describes what to look for if there could be a problem with the puncture. It is possible to use a vega xtraseal together with an xtraflange, which might give a better fixation of the device. Device is lost. Device will not be returned.
Patient Sequence No: 1, Text Type: N, H10


[56920529] Patient uses the provox vega xtraseal voice prosthesis. Complaint description from local representative: "patient was walking through his house and realized his vp is missing. Told his wife he would drive himself to the hospital. Went to the hospital and they took x-rays and those came back negative not showing anything. Gave him paperwork to call his doctor if his condition worsened and sent him home. The wife contacted the slp (b)(6), who did a house call. He used some sort of red rubber piece that he cut off of something. The wife wasn't sure what it was, but it was used to temporarily plug the puncture hole. Was told a new rx will be sent over shortly for a provox 2 4. 5mm. I asked to verify and there is no clue as to what happened, why or where the xtraseal that became dislodged is at this point in time. Note: no product will be returned. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number8032044-2016-00011
MDR Report Key6013490
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-10-10
Date of Report2016-10-10
Date of Event2016-10-02
Date Mfgr Received2016-10-03
Device Manufacturer Date2014-08-01
Date Added to Maude2016-10-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. ELIN ALGOTSSON
Manufacturer StreetKRAFTGATAN 8
Manufacturer CityHORBY, SE-24222
Manufacturer CountrySW
Manufacturer PostalSE-24222
Manufacturer Phone641519800
Manufacturer G1ATOS MEDICAL AB
Manufacturer StreetPO BOX 183 KRAFTGATAN 8
Manufacturer CityHORBY, SE-242 22
Manufacturer CountrySW
Manufacturer Postal CodeSE-242 22
Single Use3
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROVOX VEGA XTRASEAL
Generic NameVOICE PROSTHESIS
Product CodeEWL
Date Received2016-10-10
Catalog Number7782-01
Lot Number1601162
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Age11 MO
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerATOS MEDICAL AB
Manufacturer AddressPO BOX 183 KRAFTGATAN 8 HORBY, SE-242 22 SW SE-242 22


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.