MALECOT M001224110 22-411

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-10-11 for MALECOT M001224110 22-411 manufactured by Boston Scientific - Costa Rica (coyol).

Event Text Entries

[57052324] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[57052325] It was reported a catheter fracture occurred. During the nephrostomy change of a malecot drainage catheter placed in the kidney, the tube fragmented upon removal. The remainder of the tube was able to extracted. No further patient complications were reported.
Patient Sequence No: 1, Text Type: D, B5


[61410432] Device evaluated by manufacturer: a visual examination identified a foreign material attach, as part of overall visual revision. The catheter is blackened in different locations, broken and it is bent in several locations. The batch number is unknown and the manufacturing records for the complaint device could not be reviewed. The most probable root cause was unable to be determined. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[61410433] It was reported a catheter fracture occurred. During the nephrostomy change of a malecot drainage catheter placed in the kidney, the tube fragmented upon removal. The remainder of the tube was able to extracted. No further patient complications were reported.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2134265-2016-08919
MDR Report Key6018417
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-10-11
Date of Report2016-09-13
Date Mfgr Received2016-11-07
Date Added to Maude2016-10-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRADIOLOGIC TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. SONALI ARANGIL
Manufacturer StreetONE SCIMED PLACE
Manufacturer CityMAPLE GROVE MN 55311
Manufacturer CountryUS
Manufacturer Postal55311
Manufacturer Phone7634941700
Manufacturer G1BOSTON SCIENTIFIC - COSTA RICA (COYOL)
Manufacturer Street2546 FIRST STREET PROPARK FREE ZONE
Manufacturer CityALAJUELA
Manufacturer CountryCS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMALECOT
Generic NameCATHETER, PERITONEAL
Product CodeGBW
Date Received2016-10-11
Returned To Mfg2016-09-29
Model NumberM001224110
Catalog Number22-411
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - COSTA RICA (COYOL)
Manufacturer Address2546 FIRST STREET PROPARK FREE ZONE ALAJUELA CS


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-10-11