WYNMAN CARB CROWN GRIPPER 72-10TC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-10-14 for WYNMAN CARB CROWN GRIPPER 72-10TC manufactured by Integra York, Pa Inc..

Event Text Entries

[57462297] On 9/28/16 integra investigation completed. Manufacture date unknown. Method: failure analysis, device history evaluation. Results: failure analysis - a gripper in used condition, not showing any unusual markings, showing wear, staining and discoloration, battered finish, and broken insert. During the visual inspection of the instrument, it is noticed that the finish is worn and there is staining/discoloration. It is also noticed that the insert is broken on one of the tips. The damage to the insert appears to have started as a hairline fractured, leading to the breakage. The complaint report is confirmed. Device history evaluation - nonconforming product report / nonconforming material report history: none. Variance authorization / deviation history: none. Engineering change order/manufacturing change order history: there is no applicable engineering change order/manufacturing change order history. Corrective action preventive action history/corrections: none. Health hazard evaluation history: none. Conclusion: the root cause has not been identified as a workmanship or material deficiency.
Patient Sequence No: 1, Text Type: N, H10


[57462298] Customer initially reports the insert broke while using. On (b)(6) 2016 doctor reports he was gripping a crown when a tip of the carbide insert broke off in the mouth and was retrieved. No harm done.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2016-00164
MDR Report Key6029644
Report SourceUSER FACILITY
Date Received2016-10-14
Date of Report2016-09-15
Date Mfgr Received2016-09-15
Date Added to Maude2016-10-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL ASSISTANT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SANDRA LEE
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA YORK, PA INC.
Manufacturer Street589 DAVIES DRIVE
Manufacturer CityYORK PA 17402
Manufacturer CountryUS
Manufacturer Postal Code17402
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWYNMAN CARB CROWN GRIPPER
Generic NameM51 - GENERAL DENTISTRY
Product CodeEIS
Date Received2016-10-14
Returned To Mfg2016-09-26
Catalog Number72-10TC
Lot NumberUNKNOWN
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA YORK, PA INC.
Manufacturer Address589 DAVIES DRIVE 589 DAVIES DRIVE YORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-14

© 2025 FDA.report
This site is not affiliated with or endorsed by the FDA.