RENASYS GO POWER SUPPLY 66800161

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-10-17 for RENASYS GO POWER SUPPLY 66800161 manufactured by Smith & Nephew, Inc..

Event Text Entries

[57574929] (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[57574930] Arcing it was reported whilst plugging in the new power pack , one of our pnps received an electric shock. When the unit was returned to the rep she also tried to plug the unit and received an electric shock.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3006760724-2016-00156
MDR Report Key6033029
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-10-17
Date of Report2017-07-14
Date of Event2016-10-06
Date Mfgr Received2016-10-10
Date Added to Maude2016-10-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMRS CLAUDIA ODOY
Manufacturer StreetSCHACHENALLEE 29
Manufacturer CityAARAU 5001
Manufacturer CountrySZ
Manufacturer Postal5001
Manufacturer Phone0628320660
Manufacturer G1SMITH & NEPHEW ENDOSCOPY
Manufacturer Street76. S MERIDAN AVENUE
Manufacturer CityOKLAHOMA CITY OK 73107
Manufacturer CountryUS
Manufacturer Postal Code73107
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRENASYS GO POWER SUPPLY
Generic NamePUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)
Product CodeBTA
Date Received2016-10-17
Model Number66800161
Catalog Number66800161
Lot Number144701649
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSMITH & NEPHEW, INC.
Manufacturer Address970 LAKE CARILLON DRIVE ST PETERSBURG FL 33716 US 33716


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-17