PRECISION MEDICAL INC. PM65

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a health professional,user faci report with the FDA on 2016-10-17 for PRECISION MEDICAL INC. PM65 manufactured by Precision Medical Inc..

Event Text Entries

[57687152] (b)(6). Subject: easy go aspirator serial # (b)(4). Med watch (b)(4). Dear (b)(4), the pm65 serial number (b)(4), complaint of getting hot and smoking. The history on the device is that it was manufactured in december of 1999, and precision medical shows no record of the device being returned for repair. The device is 16 years old, or three times the recorded life of the product. The likely cause of the complaint, is that the lead acid battery contained in the device, over heated, when the fluid leaked out. This caused the smell and smoke. I would recommend that all pm65 aspirators older than 5 years, be serviced, and at the very least the batteries should be replaced if older than 2 years. The life of the pm65 aspirator is 5 years. The warranty of the pm65 is 2 years. The life cycle of a lead acid battery is approximately 2 years. I would also recommend that pm65 devices, three times the average life span, be considered for replacement. Any other questions please feel free to contact me. Regards, (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[57687153] Event desc: the unit began to smoke and create an acid bad odor and the unit was extremely hot to touch. This incident occurred at approximately 4:40 pm and even after 15 minutes the unit was still warm to touch. After being persuasive, i was able to get the battery out of the unit ; on further inspection, i noticed that the charging circuit board was burnt so it must have shorted out. Honestly because the case is melted and distorted and the charging circuit the unit is compromised, i believe that this unit is beyond repair and should be replaced. I will call the manufacturer and see if there has been any instances of this happening and i will also try to get a price on a replacement unit.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523148-2016-00003
MDR Report Key6034053
Report SourceHEALTH PROFESSIONAL,USER FACI
Date Received2016-10-17
Date of Report2016-07-22
Date of Event2016-06-30
Report Date2016-07-01
Date Reported to FDA2016-07-01
Date Mfgr Received2016-08-19
Device Manufacturer Date1999-12-01
Date Added to Maude2016-10-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR JAMES PARKER
Manufacturer Street300 HELD DRIVE
Manufacturer CityNORTHAMPTON PA 18067
Manufacturer CountryUS
Manufacturer Postal18067
Manufacturer Phone6102626090
Manufacturer G1PRECISION MEDICAL INC.
Manufacturer Street300 HELD DRIVE
Manufacturer CityNORTHAMPTON 18067
Manufacturer CountryUS
Manufacturer Postal Code18067
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRECISION MEDICAL INC.
Generic NamePM65 EASY GO VAC
Product CodeBTA
Date Received2016-10-17
Model NumberPM65
Catalog NumberPM65
Lot NumberNA
OperatorBIOMEDICAL ENGINEER
Device AvailabilityY
Device Age16 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerPRECISION MEDICAL INC.
Manufacturer Address300 HELD DR NORTHAMPTON PA 18067 US 18067


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-10-17

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.