ADU *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2005-05-06 for ADU * manufactured by Datex Ohmeda.

Event Text Entries

[380487] Both of the ups on the datex ohmeda adu anesthesia machine failed and shut down the machine. The batteries failed and datex ohmeda replaced both of the ups'. The ups is used for surge protection and should be removed from the units or replaced with a more satisfactory model. The 2 ups supplied by datex ohmeda are manufactured by oneac.
Patient Sequence No: 1, Text Type: D, B5


[425887] Per your letter dated 5/31/2005 requesting a faiure analysis of subject report, datex-ohmeda has received a response from oneac, the mfr of the ups unit involved in the alleged event. Their response is quoted as follows: "the ups used by datex-ohmeda does the following: 1. Is a medical grade ups offering low electrical cumulative leakage for the ups and connected equipment of less than 300 microamps 2. Provide surge protection for connected equipment 3. Battery backup in high line, low line, and no power. 4. The ups alarms when it goes to battery having received some of the upss back from the hosp and after talking to datex-ohmeda personnel, oneac believes the following occurred and would like this added to your medwatch response: 'the onm600sj-si functioned by going to battery back up under a faulty mains condition, thus keeping the attached equipment functional under these conditions. Upon depletion of the batteries, with faulty main conditions still present, the ups turned off, either due to complete loss of mains, or to protect the attached equipment from the faulty mains condition. The ups functioned properly in these actions. '"
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number603605
MDR Report Key603605
Date Received2005-05-06
Date of Report2005-05-06
Date of Event2005-05-05
Report Date2005-05-06
Date Reported to FDA2005-05-06
Date Added to Maude2005-05-19
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameADU
Generic NameANESTHESIA MACHINE
Product CodeBSZ
Date Received2005-05-06
Model NumberADU
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityR
Device Age6 MO
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key593441
ManufacturerDATEX OHMEDA
Manufacturer Address* * *

Device Sequence Number: 2

Brand NameONEAC
Generic NameUNINTERRUPTIBLE POWER SUPPLY, UPS
Product CodeMPE
Date Received2005-05-06
Model Number*
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityR
Device Age6 MO
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key593443
ManufacturerONEAC
Manufacturer Address27944 NORTH BRADLEY ROAD LIBERTYVILLE IL 600489700 US


Patients

Patient NumberTreatmentOutcomeDate
10 2005-05-06

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.