UNIVER TOFFLEMIRE TYPE RE 72-34

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a user facility report with the FDA on 2016-10-18 for UNIVER TOFFLEMIRE TYPE RE 72-34 manufactured by Integra York, Pa Inc..

Event Text Entries

[57722444] Integra has completed their internal investigation on october 12, 2016. The investigation included: methods: evaluation of actual device. Review of device history records. Review of complaints history. Results: evaluation of returned device; there were ten tofflemire retainers returned with no unusual markings and complaint stating that the doctor cut himself. Upon further investigation it is found that most do have sharp edges on them. Dhr review; there is no applicable nonconforming product report / nonconforming material report history complaints history; complaint metrics tracked, trended, and reviewed monthly with management. Conclusion: the complaint report is confirmed; the root cause has been determined to be a workmanship or material deficiency. Appropriate action has been implemented to rectify this manufacturing deficiency.
Patient Sequence No: 1, Text Type: N, H10


[57722445] Customer initially reports device defective cuts doctors hand. On 9/30/2016 doctor reports she was applying a band to a lower r back molar when she was cut on sharp threads of the device. No special treatment necessary, no infection. No harm to patient
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2523190-2016-00168
MDR Report Key6036495
Report SourceUSER FACILITY
Date Received2016-10-18
Date of Report2016-09-26
Date Mfgr Received2016-09-26
Date Added to Maude2016-10-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactUSER SANDRA LEE
Manufacturer Street311 ENTERPRISE DRIVE
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362393
Manufacturer G1INTEGRA YORK, PA INC.
Manufacturer Street589 DAVIES DRIVE
Manufacturer CityYORK PA 17402
Manufacturer CountryUS
Manufacturer Postal Code17402
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNIVER TOFFLEMIRE TYPE RE
Generic NameM51 - GENERAL DENTISTRY
Product CodeJEP
Date Received2016-10-18
Returned To Mfg2016-10-12
Catalog Number72-34
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA YORK, PA INC.
Manufacturer Address589 DAVIES DRIVE 589 DAVIES DRIVE YORK PA 17402 US 17402


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-18

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.