CAPILLARY ADAPTERS 906-018 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2005-05-16 for CAPILLARY ADAPTERS 906-018 * manufactured by Radiometer Medical Aps.

Event Text Entries

[387850] A small plastic adapter is necessary in order to transfer a blood sample from a capillary tube to a blood gas analyzer type abl5. Following a transfer the user noticed that the po2 values were too high and approaching that of ambient air. A bubble was subsequently discovered in the measuring chamber. Othe tests revealed that the introduction of ambient air is possible when using the adapter. No pt was maltreated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3002807968-2005-00004
MDR Report Key603669
Report Source01,05,06
Date Received2005-05-16
Date of Report2005-05-12
Date of Event2005-03-04
Date Mfgr Received2005-03-15
Device Manufacturer Date2005-01-01
Date Added to Maude2005-05-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactLEIF GUDNITZ
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Manufacturer Phone8273228
Manufacturer G1RADIOMETER MEDICAL APS
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ
Manufacturer CountryDA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAPILLARY ADAPTERS
Generic NameADAPTOR FOR CONNECTING CAPILLARY TUBES
Product CodeGIO
Date Received2005-05-16
Model Number906-018
Catalog Number*
Lot NumberALL 2005
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key593512
ManufacturerRADIOMETER MEDICAL APS
Manufacturer Address* * *
Baseline Brand NameCAPILLARY ADAPTERS
Baseline Generic NameADAPTOR FOR CONNECTING CAPILLARY TUBES
Baseline Model No906-018
Baseline Catalog No*
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2005-05-16

© 2026 FDA.report