TRIAGE TOTAL 5 CONTROL L1 88753

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-10-18 for TRIAGE TOTAL 5 CONTROL L1 88753 manufactured by .

Event Text Entries

[57804984] Mdr is being filed due to field action (b)(4). Customer's complaint was replicated with in-house testing of the code chip of retain lot c3233a. Code chip ranges did not match ev card ranges. The manufacturing records for the control lot were reviewed and the lot met release specifications. A capa, capa-(b)(4), was initiated to address this issue.
Patient Sequence No: 1, Text Type: N, H10


[57805025] The support consultant attempted to run the total 5 control level 1 twice at 2 stdev with myoglobin results outside of the expected range high at 132 ng/ml and 131 ng/ml (2 stdev: 71. 9-119 ng/ml). The support consultant installed a code chip from a different box of controls with the same lot number and then attempted to run the total 5 control level 1 at 3 stdev with a myoglobin result outside of the expected range high at 124 ng/ml (3 stdev: 60. 2-131 ng/ml).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2027969-2016-00653
MDR Report Key6038570
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-10-18
Date of Report2016-09-29
Date of Event2016-09-09
Date Mfgr Received2016-09-29
Date Added to Maude2016-10-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactYA-LING KING
Manufacturer Street9975 SUMMERS RIDGE ROAD
Manufacturer CitySAN DIEGO CA 92121
Manufacturer CountryUS
Manufacturer Postal92121
Manufacturer Phone8588052084
Single Use3
Remedial ActionRL
Previous Use Code3
Removal Correction Number2027969-10/12/16-004-C
Event Type3
Type of Report3

Device Details

Brand NameTRIAGE TOTAL 5 CONTROL L1
Generic NameMULTI-ANALYTE CONTROLS
Product CodeJJY
Date Received2016-10-18
Returned To Mfg2016-10-11
Model Number88753
Lot NumberC3233A
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-18

© 2024 FDA.report