MERGE EYE STATION MERGE EYE STATION V10.6.46

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2016-10-19 for MERGE EYE STATION MERGE EYE STATION V10.6.46 manufactured by Merge Healthcare.

Event Text Entries

[57807950] Merge healthcare is investigating the issue. Additional information has been requested from the customer but has not yet been received.
Patient Sequence No: 1, Text Type: N, H10


[57807951] Merge eye station is intended to be used in conjunction with existing ophthalmic fundus cameras to take images of the eye, perform fluorescein angiography, red free, color and icg still-image photography as well as video imaging. On (b)(6) 2016, merge healthcare was notified that an existing eye station was failing and a replacement system was required. Merge support shipped new cables but the issue was not resolved. On 10/04/2016, information was received from the account that indicated the issue resulted in patients being rescheduled. The rescheduling of patients resulted in a delay in diagnosis and/or treatment. Additional information has been requested but has not yet been received. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2183926-2016-00763
MDR Report Key6039025
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2016-10-19
Date of Report2016-09-05
Date of Event2016-09-05
Date Mfgr Received2016-10-04
Date Added to Maude2016-10-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. MICHAEL DIEDRICK
Manufacturer Street900 WALNUT RIDGE DRIVE
Manufacturer CityHARTLAND, WI 53029
Manufacturer CountryUS
Manufacturer Postal53029
Manufacturer Phone2629123570
Manufacturer G1MERGE HEALTHCARE
Manufacturer Street900 WALNUT RIDGE DR.
Manufacturer CityHARTLAND, WI 53029
Manufacturer CountryUS
Manufacturer Postal Code53029
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMERGE EYE STATION
Generic NameCAMERA, OPHTHALMIC, AC-POWERED
Product CodeHKI
Date Received2016-10-19
Model NumberMERGE EYE STATION V10.6.46
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMERGE HEALTHCARE
Manufacturer Address900 WALNUT RIDGE DRIVE HARTLAND, WI 53029 US 53029


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.