3M TRUE DEFINITION SCANNER 78007

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,foreig report with the FDA on 2016-10-19 for 3M TRUE DEFINITION SCANNER 78007 manufactured by 3m Espe Dental Products.

Event Text Entries

[57926605] An in office electrical evaluation was conducted and found that the outlet in the dental office was non-conforming, as it was not appropriately gounded for the connection to the 3m true definition scanner. Appropriate electrical safety measures (isolation transformer) are provided with the true definition scanner system. The 3m true definition scanner instructions for use does contain the warning, 'to reduce the risks associated with hazardous voltage and fire; use only a properly grounded power outlet; do not use extension cords or multiple portable power socket outlets.
Patient Sequence No: 1, Text Type: N, H10


[57926606] On (b)(6) 2016, it was reported that a female patient in (b)(6) experienced an electrical impulse during use of the 3m true definition scanner.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005174370-2016-00112
MDR Report Key6042703
Report SourceCOMPANY REPRESENTATIVE,FOREIG
Date Received2016-10-19
Date of Report2016-09-22
Date of Event2016-09-20
Date Mfgr Received2015-09-22
Date Added to Maude2016-10-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactANGIE DRAPER
Manufacturer Street2510 CONWAY AVENUE
Manufacturer CityST. PAUL MN 551441000
Manufacturer CountryUS
Manufacturer Postal551441000
Manufacturer Phone6517331179
Manufacturer G13M ESPE DENTAL PRODUCTS
Manufacturer Street2510 CONWAY AVENUE
Manufacturer CityST. PAUL MN 551441000
Manufacturer CountryUS
Manufacturer Postal Code551441000
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3M TRUE DEFINITION SCANNER
Generic NameSYSTEM,OPTICAL IMPRESSION, COMPUTER ASSISTED DESIGN AND MANUFACTURING (CAD/CAM)
Product CodeNOF
Date Received2016-10-19
Catalog Number78007
OperatorDENTIST
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
Manufacturer3M ESPE DENTAL PRODUCTS
Manufacturer Address2510 CONWAY AVENUE ST. PAUL MN 551441000 US 551441000


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-10-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.