BIPASS DISPOSABLE PUSHER SINGLE PACK N/A 902093

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-10-20 for BIPASS DISPOSABLE PUSHER SINGLE PACK N/A 902093 manufactured by Biomet Sports Medicine.

Event Text Entries

[57984511] Current information is insufficient to permit a conclusion as to the cause of the event. Review of device history records show the lot released with no recorded anomaly or deviation. There are warnings in the package insert that state that this type of event can occur: under warnings, number 4 states, "do not apply excessive force to the pusher actuator. Repeat step 3a if necessary to avoid excessive force. "
Patient Sequence No: 1, Text Type: N, H10


[57984512] During a rotator cuff repair the suture wire kinked and bent while trying to use and would not pass suture as intended. A second wire was used with minimal delay and no serious injury.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0001825034-2016-04203
MDR Report Key6044955
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-10-20
Date of Report2016-09-21
Date of Event2016-09-20
Date Mfgr Received2016-09-21
Device Manufacturer Date2015-10-30
Date Added to Maude2016-10-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. MEGAN HAAS
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal46582
Manufacturer Phone5743726700
Manufacturer G1BIOMET ORTHOPEDICS
Manufacturer Street56 E. BELL DRIVE
Manufacturer CityWARSAW IN 46582
Manufacturer CountryUS
Manufacturer Postal Code46582
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameBIPASS DISPOSABLE PUSHER SINGLE PACK
Generic NameORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Product CodeHXO
Date Received2016-10-20
Model NumberN/A
Catalog Number902093
Lot Number869410
ID NumberN/A
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBIOMET SPORTS MEDICINE
Manufacturer Address56 E. BELL DRIVE WARSAW IN 46582 US 46582


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-20

© 2026 FDA.report