UPSYLON? Y MESH KIT M0068318220

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,health report with the FDA on 2016-10-20 for UPSYLON? Y MESH KIT M0068318220 manufactured by Boston Scientific - Marlborough.

Event Text Entries

[58017739] The exact age of the patient is unknown, however, it was reported the patient was over 18 years. The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown. However, the complainant reported that the device was not expired. The complainant indicated that the device is implanted and will not be returned for evaluation; therefore a failure analysis of the complaint device could not be completed. If any further relevant information is identified, a supplemental medwatch will be filed.
Patient Sequence No: 1, Text Type: N, H10


[58017740] It was reported to boston scientific corporation that an upsylon y-mesh kit was implanted during a lap assisted sacrocolpopexy (non-robotic) procedure. According to the complainant, 3 months post procedure, the patient returned with recurrent prolapse and it was found that the sacral piece of the mesh tore off from the anterior and posterior pieces at the connection point. Revision was done using a polyform synthetic mesh in (b)(6) 2016 with no issues, and the patient is reported to be fine.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005099803-2016-03095
MDR Report Key6045962
Report SourceCOMPANY REPRESENTATIVE,HEALTH
Date Received2016-10-20
Date of Report2016-09-26
Date Mfgr Received2016-09-26
Date Added to Maude2016-10-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactEMP. NANCY CUTINO
Manufacturer Street100 BOSTON SCIENTIFIC WAY
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5086834000
Manufacturer G1PROXY BIOMEDICAL LIMITED
Manufacturer StreetCOILLEACH SPIDDAL
Manufacturer CityGALWAY
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUPSYLON? Y MESH KIT
Generic NameGYNECOLOGIC LAPAROSCOPE AND ACCESSORIES
Product CodeOHD
Date Received2016-10-20
Model NumberM0068318220
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerBOSTON SCIENTIFIC - MARLBOROUGH
Manufacturer Address100 BOSTON SCIENTIFIC WAY MARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2016-10-20

© 2026 FDA.report