HARMONY IQ INTEGRATION SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-10-21 for HARMONY IQ INTEGRATION SYSTEM manufactured by Steris Corporation - Vts Medical Systems.

Event Text Entries

[58508683] During the procedure the image displayed on the monitors in the surgical room was not available. The harmony iq integration system required a restart subsequently causing a delay in the procedure. After the image was restored the patient procedure was completed successfully. No report of injury. A steris service technician inspected the harmony iq integration system and found the video signal to the monitors required re-routing. The technician tested the system and confirmed it to be operating properly. No additional issues have been reported.
Patient Sequence No: 1, Text Type: N, H10


[58508684] The user facility reported that during a patient procedure their harmony iq integration system did not operate properly.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000404456-2016-00009
MDR Report Key6048106
Date Received2016-10-21
Date of Report2016-10-21
Date of Event2016-09-23
Date Mfgr Received2016-09-23
Date Added to Maude2016-10-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CORPORATION - VTS MEDICAL SYSTEMS
Manufacturer Street40 MELVILLE PARK RD
Manufacturer CityMELVILLE NY 11747
Manufacturer CountryUS
Manufacturer Postal Code11747
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand NameHARMONY IQ INTEGRATION SYSTEM
Generic NameINTEGRATION SYSTEM
Product CodeKQM
Date Received2016-10-21
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - VTS MEDICAL SYSTEMS
Manufacturer Address40 MELVILLE PARK RD MELVILLE NY 11747 US 11747


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-21