CRYOSTAT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-10-26 for CRYOSTAT manufactured by Leica Biosystems Richmond.

Event Text Entries

[58380254]
Patient Sequence No: 1, Text Type: N, H10


[58380255] While cutting patient tissue the cryostat began "chunking out" or cutting too deeply and too quickly into the specimen, thereby leaving the tissue unreadable and not viable. The provider was immediately notified and the "troubleshooting" steps at the operator level were performed. These steps were deemed not helpful in rectifying the problem, again the provider was notified and leica, the manufacturer was notified. Leica arrived on site and substantiated, that indeed the machine was not performing correctly and was giving inconsistent results and began "troubleshooting" at the manufacturer level. After several failed attempts to correct the machine, it was determined that a new head was to be installed in the machine the next day and that a "loaner" machine would be sent overnight for the next business day. Unfortunately, during the machine malfunction, the patient's specimen was damaged and not able to be interpreted. This required additional tissue to be taken and sent for permanent section. Manufacturer response for cryostat, (brand not provided) (per site reporter): refer to description of event.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number6056770
MDR Report Key6056770
Date Received2016-10-26
Date of Report2016-09-23
Date of Event2016-08-03
Report Date2016-09-23
Date Reported to FDA2016-09-23
Date Reported to Mfgr2016-09-23
Date Added to Maude2016-10-26
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPATIENT
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCRYOSTAT
Generic NameMICROTOME, CRYOSTAT
Product CodeIDL
Date Received2016-10-26
Lot Number1510S-3
ID Number347302
Device Availability*
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerLEICA BIOSYSTEMS RICHMOND
Manufacturer Address5205 US HIGHWAY 12 RICHMOND IL 60071 US 60071


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.