CLINIMACS CD34 REAGENT N/A 171-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative report with the FDA on 2016-10-27 for CLINIMACS CD34 REAGENT N/A 171-01 manufactured by Miltenyi Biotec Gmbh.

Event Text Entries

[58515405] The customer did not provide any data for a comprehensive investigation.
Patient Sequence No: 1, Text Type: N, H10


[58515406] The customer performed a clinimacs cd34 enrichment procedure on the clinimacs plus instrument using the clinimacs tubing set. After the enrichment procedure, the customer reported that approx. 40% of the cd34 antigen positive stem cells were found in the negative fraction bag.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005290010-2016-00095
MDR Report Key6060381
Report SourceCOMPANY REPRESENTATIVE
Date Received2016-10-27
Date of Report2016-10-26
Date of Event2016-09-23
Date Mfgr Received2016-09-28
Device Manufacturer Date2016-08-18
Date Added to Maude2016-10-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS NANCY JOHANSEN
Manufacturer Street85 HAMILTON ST
Manufacturer CityCAMBRIDGE MA 02139
Manufacturer CountryUS
Manufacturer Postal02139
Manufacturer Phone6172180062
Manufacturer G1MILTENYI BIOTEC GMBH
Manufacturer StreetFRIEDRICH-EBERT-STRASSE 68
Manufacturer CityBERGISCH GLADBACH, 51429
Manufacturer CountryGM
Manufacturer Postal Code51429
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINIMACS CD34 REAGENT
Generic NameCLINIMACS CD34 REAGENT
Product CodeOVG
Date Received2016-10-27
Model NumberN/A
Catalog Number171-01
Lot Number6160811005
Device Expiration Date2017-01-17
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMILTENYI BIOTEC GMBH
Manufacturer AddressFRIEDRICH-EBERT-STRASSE 68 BERGISCH GLADBACH, 51429 GM 51429


Patients

Patient NumberTreatmentOutcomeDate
10 2016-10-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.