BN II / BN PROSPEC? SYSTEM OSAR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a foreign,health professional,u report with the FDA on 2016-11-02 for BN II / BN PROSPEC? SYSTEM OSAR manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[58935545] The cause of the falsely elevated igg results is unknown. Siemens healthcare headquarters support center and technical support laboratory representatives evaluated the data available. No systemic instrument or reagent issues were identified. The device is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


[58935546] Falsely elevated immunoglobulin g (igg) results were obtained on two patient samples on the behring nephelometer ii (bn-ii) instrument. The patient results were not reported to the physician because the values exceeded the sum of the igg subclass determinations. The repeat test of the same samples on an alternate bnii instrument gave lower results consistent with the sum of the subclasses. The repeated results were reported to the physician. There is no indication of adverse impact to the patients due to the falsely elevated igg results, which were not reported to the physician.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9610806-2016-00030
MDR Report Key6072324
Report SourceFOREIGN,HEALTH PROFESSIONAL,U
Date Received2016-11-02
Date of Report2016-11-02
Date of Event2016-09-01
Date Mfgr Received2016-10-03
Device Manufacturer Date2016-05-11
Date Added to Maude2016-11-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Street76 EMIL VON BEHRING STRASSE
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBN II / BN PROSPEC? SYSTEM
Generic NameN ANTISERA TO HUMAN IMMUNOGLOBULINS (IGG, IGA AND IGM)
Product CodeCFN
Date Received2016-11-02
Catalog NumberOSAR
Lot Number153022D
Device Expiration Date2018-04-26
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer AddressEMIL VON BEHRING STRASSE 76 MARBURG, D-35041 GM D-35041


Patients

Patient NumberTreatmentOutcomeDate
10 2016-11-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.