120 ML BD VACUTAINER? PLASTIC URINE COLLECTION CUP WITH SAMPLING DEVICE 364975

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2016-11-03 for 120 ML BD VACUTAINER? PLASTIC URINE COLLECTION CUP WITH SAMPLING DEVICE 364975 manufactured by Becton, Dickinson & Co..

Event Text Entries

[59455753] Medical device brand name: 120 ml bd vacutainer? Plastic urine collection cup with integrated sampling device. (b)(4). Device evaluation: result - no samples or photos were returned for evaluation. A review of the device history record cannot be completed as the lot number was not provided for this incident. Conclusion - although based on receipt of this complaint bd acknowledges that the customer has had an issue with this particular product, for plant evaluation purposes this is not confirmed. Bd will continue to monitor this defect for tracking and trending purposes. The bd quality engineer notes that the product label reads "caution: product contains needle under label".
Patient Sequence No: 1, Text Type: N, H10


[59455754] It was reported that elderly patients with poor eyesight are removing the yellow sticker covering the needle on the suspect device and sticking themselves on the unused needle. This has been reported to have occurred five times.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1917413-2016-00016
MDR Report Key6077438
Date Received2016-11-03
Date of Report2016-10-19
Date of Event2016-10-14
Date Mfgr Received2016-10-14
Date Added to Maude2016-11-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactBDX BRETT WILKO
Manufacturer Street9450 SOUTH STATE STREET
Manufacturer CitySANDY UT 84070
Manufacturer CountryUS
Manufacturer Postal84070
Manufacturer Phone8015652845
Manufacturer G1BECTON, DICKINSON & CO.
Manufacturer Street150 SOUTH 1ST AVENUE
Manufacturer CityBROKEN BOW NE 68822
Manufacturer CountryUS
Manufacturer Postal Code68822
Single Use3
Previous Use Code3
Event Type3
Type of Report0

Device Details

Brand Name120 ML BD VACUTAINER? PLASTIC URINE COLLECTION CUP WITH SAMPLING DEVICE
Generic NameCOLLECTION CUP
Product CodeJXA
Date Received2016-11-03
Catalog Number364975
Lot NumberUNKNOWN
OperatorLAY USER/PATIENT
Device Availability*
Device Eval'ed by MfgrR
Device Sequence No0
Device Event Key0
ManufacturerBECTON, DICKINSON & CO.
Manufacturer Address150 SOUTH 1ST AVENUE BROKEN BOW NE 68822 US 68822


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2016-11-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.