MULTISTIX 10-SG 100 USA 10336425

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a company representative,user f report with the FDA on 2016-11-07 for MULTISTIX 10-SG 100 USA 10336425 manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[59353281] The customer states she has prescribed antibiotics to patients with elevated results and has ordered further testing. The customer has received replacement strips and is fully operational on the new strips.
Patient Sequence No: 1, Text Type: N, H10


[59353282] The customer reported that some leukocyte results were incorrectly reported out due to deteriorated strips and patients were prescribed antibiotics that were not necessary. Specific data was not provided on these samples.
Patient Sequence No: 1, Text Type: D, B5


[63925760] The reagent was returned for investigation. The investigation of the desiccants demonstrated that the strips were exposed to excessive moisture. Visual appearance of strips also consistent with moisture exposure. Per the multistix ifu: "protection against exposure to light, heat, and ambient moisture is mandatory to guard against altered reagent reactivity". "discoloration or darkening of the test pads may indicate deterioration. " the customer is operational.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002637618-2016-00138
MDR Report Key6082693
Report SourceCOMPANY REPRESENTATIVE,USER F
Date Received2016-11-07
Date of Report2016-12-12
Date of Event2016-10-11
Date Mfgr Received2016-11-21
Date Added to Maude2016-11-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVEN ANDBERG
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693655
Manufacturer G1KIMBALL ELECTRONICS
Manufacturer StreetUL. POZENANSKA 1/C POLAND SP.Z O.O
Manufacturer CityTARNOWO PODGORNE, 62080
Manufacturer CountryPL
Manufacturer Postal Code62080
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMULTISTIX 10-SG 100 USA
Generic NameMULTISTIX 10 SG
Product CodeJIR
Date Received2016-11-07
Catalog Number10336425
Lot Number512065
Device Expiration Date2017-06-30
OperatorPHYSICIAN
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591