N
Patient 1
THE CAUSE OF THE DISCREPANT FALSELY DEPRESSED D-DIMER RESULT IS UNKNOWN. HOWEVER, ANALYSIS OF THE INSTRUMENT DATA REVEALED NO SYSTEMATIC FAILURES OR QC FAILURES. THE ISSUE IS AN ISOLATED OBSERVATION OF A PATIENT MEASUREMENT WITH A FALSELY LOW D-DIMER RESULT IN A SINGLE DETERMINATION WHICH IS DISCORDANT TO OTHER MEASUREMENTS. THE CAUSE OF THE ISSUE SEEMS TO BE RELATED TO THE AFFECTED 2ND SAMPLE. THIS SAMPLE IS DESCRIBED AS HAVING A HIGH COUNT OF FIBRIN MONOMERS. FIBRIN MONOMERS ARE AN INTERMEDIATES IN THE FIBRINOGEN ACTIVATION TO FIBRIN AND MAY CAUSE MICROCLOTS. SAMPLES WITH MICROCLOTS MAY CAUSE SAMPLE UNDER-ASPIRATION BY PIPETTOR CLOGGING AND BY THIS MAY DELIVER FALSE LOW RESULTS. MICROCLOTS CAN BE REMOVED BY CORRECT SAMPLE PREPARATION WITH CENTRIFUGATION, AND THEREFORE THE PRESENCE OF MICROCLOTS IS A PREANALYTICAL ISSUE WHICH IS UNDER THE USER'S RESPONSIBILITY. IN ADDITION, AS THE PATIENT HAS A KNOWN MALIGNANCY SUSPECTED FOR DIC/SEPSIS, THE RESULTS OF SUCH SAMPLES HAVE TO BE INTERPRETED WITH CAUTION AS THESE DISEASE STAGES ARE KNOWN TO POTENTIALLY INTERFERE WITH D-DIMER ANALYSIS AS INDICATED IN THE 'SUMMARY AND EXPLANATION' SECTION OF THE INSTRUCTIONS FOR USE. THE DEVICE IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.